• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM, INC. AFFINITY 4 BED; BIRTHING BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM, INC. AFFINITY 4 BED; BIRTHING BED Back to Search Results
Model Number 3700
Device Problem Device Alarm System (1012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/30/2015
Event Type  malfunction  
Event Description
Hill-rom received a report from the account stating the nurse call was inoperative.The bed was located at the account.There was no pt/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Manufacturer Narrative
The hill-rom technician found a bent pin on the communication cable.Per the hill-rom service manual the affinity three birthing bed and affinity four birthing beds require an effective maintenance program.We recommend that you perform semiannual preventive maintenance.Check the switch in the side rails to ensure they are functioning correctly.Also check for intermittent operation.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unk if the facility performs preventative maintenance on their beds.The technician repaired the bent pin to resolve the issue.Based on this info, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED
Type of Device
BIRTHING BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
john cummings
1069 state route 46 east
batesville, IN 47006
8129312869
MDR Report Key4950333
MDR Text Key22152173
Report Number1824206-2015-00745
Device Sequence Number1
Product Code HDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Unknown
Type of Report Initial
Report Date 06/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3700
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received06/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-