The hospital reported that, during a case, the agent readings were lower than dial settings, and the patient's vitals indicated they were light.The clinician switched to desflurane, and the patient's vitals settled into expected ranges.There was no reported patient injury.
|
The unit was returned to the manufacturing site for investigation.The unit was tested and the output was found low to manufacturing specifications.Further inspection revealed loose actuator nuts, resulting in the actuator pin being out of adjustment.As a result, the vaporizer manifold port valves were not depressed sufficiently to allow expected flow of anesthetic agent into the fresh gas flow stream.
|