Catalog Number 000000000000080300 |
Device Problem
Programming Issue (3014)
|
Patient Problem
No Information (3190)
|
Event Date 07/06/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation: the run data file (rdf) was analyzed for this event.The analysis of the run data file did not find a conclusive cause for the higher than expected wbc content in the platelet product reported for this collection.No unusual process variable was identified and the signals in the run data file indicate that the trima accel system operated as intended.Based on the available information, it cannot be ruled out that the higher than expected wbc content in the platelet product could be donor-related.It also cannot be ruled out that a sampling, calculation, or other process error could have contributed to the higher than expected wbc content in the platelet product.Investigation is in process.A follow-up report will be provided.
|
|
Event Description
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event.Donor unit # (b)(4).The disposable set is not available for return because it was discarded by the customer.
|
|
Manufacturer Narrative
|
Investigation: the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Root cause: this disposable set was unavailable for specific root cause analysis.Signals in the run data file do not indicate a conclusive cause for the greater than expected rwbc content in the platelet product reported for this collection.Possible root causes were provided in the initial report for this event.
|
|
Search Alerts/Recalls
|