• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE Back to Search Results
Model Number MODEL 250
Device Problem Operating System Becomes Nonfunctional (2996)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/21/2015
Event Type  malfunction  
Event Description
A user reported observing a frozen screen on her handheld device when trying to interrogate a patient's device.Instructions on how to perform a hard reset of the handheld device were provided but it is unknown if the user has attempted a hard reset of the handheld device.The handheld device and software flashcard have not been returned to date.
 
Event Description
Additional information received indicates that troubleshooting was performed and the unit was found to be working properly.It was determined that the user did not wait long enough after interrogation for the post-interrogation screen to be seen.The handheld device along with associated software, programming wand and cables were returned for product analysis.Analysis of the handheld device, software, programming wand and cables found no anomalies; all returned devices performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4960893
MDR Text Key23342801
Report Number1644487-2015-05395
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Report Date 07/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1073773
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2015
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received08/04/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/10/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age18 YR
-
-