The patient was implanted with a left ventricular assist device (lvad).It was reported that the patient was admitted to the hospital due to volume overload.It was reported that the patient had progressive deterioration of his percutaneous lead (lead), but was unable to keep his appointments with the clinic.The system controller data log file was submitted to the manufacturer for evaluation along with pictures and x-ray images of the lead.The analysis of the log file revealed a few transient power level increases.The pictures of the lead were examined and showed a large amount of tape of various types covering the lead.The x-ray images showed abnormal kinking of the lead.No abnormal pump operation was observed at the time of the analysis and a distal-end replacement of the lead was scheduled to prevent future failure.During the distal-end lead replacement, the evaluation of the lead revealed a slit in the braided shield near the lead¿s exit site from the patient.The observed compromise to the shield appeared to affect its ability to sustain a ground.This appeared to be the most likely reason why the patient did not receive any alarms.The distal-end replacement of the lead was completed successfully and subsequent issues have been reported.
|
The evaluation of the returned percutaneous lead (lead) confirmed the report of superficial lead damage.Approximately 17 inches of the external portion/distal end of the lead was returned for evaluation.Large amounts of tape were applied to the entire length of the lead.The tape and reinforcing sleeve were removed, and examination of the shielding and bionate layers revealed areas with shield breakdown, and multiple kinks and twists.Continuity testing was performed on the returned portion of the lead and no issues were found.The shielding and bionate layers were removed and examination of the underlying wires revealed disruptions in the insulation of the green wire approximately 3 inches from the metal system controller connector, and in the insulation of the black wire, approximately 10.5 inches from the metal system controller connector.As a result, the conductors of these wires were exposed.The disruptions of the wires appeared to be the result of mechanical damage, as a result of the large amounts of kinking and twisting.The disruptions in the wires could have interrupted function as a result of the exposed conductors of either the green or black wires potentially contacting the braided shield and creating an electrical short to ground if the device was supported by the power module.The electrical short to ground would have caused alarms and low speed events; however, due to the kinking and twisting of the lead, the braided shielding had deteriorated in some areas, and there might not have been a connection to sustain a ground to cause any alarm.A review of the device history records revealed no deviations from manufacturing or quality assurance specifications.The manufacturer is closing the file on this event.
|