CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/15/2015
Event Type  malfunction  
Event Description
It was reported that the handheld screen was scrambled and unable to read.The suspect handheld device has not been returned to date.
 
Event Description
The handheld device and software flashcard have been returned to the manufacturer where analysis is currently underway.
 
Event Description
Analysis of the returned handheld device determined that the cause for the reported event was associated with a defective display.Once the display was replaced with a known good display, no further anomalies associated with the handheld were noted during testing using the ac adapter or the main battery with a full charge.No anomalies associated with flashcard software were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4964324
MDR Text Key23441943
Report Number1644487-2015-05439
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Followup
Report Date 07/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1067570
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received09/08/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/12/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-