MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C4471
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/18/2015
Event Type  malfunction  
Event Description
It was reported that the patient felt an electric shock from the homechoice device during use.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).The device was received and an evaluation was performed to investigate the reported event.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.The device was determined to meet functional and electrical performance specification requirements per rite testing.In addition, a visual inspection was performed.The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Upon conclusion of the investigation, the reported issue could not be verified.Should additional relevant additional information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4965377
MDR Text Key18995514
Report Number1416980-2015-31288
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Company Representative,company representati
Reporter Occupation Other
Type of Report Initial
Report Date 09/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C4471
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received09/16/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-