(b)(4).The device was received and an evaluation was performed to investigate the reported event.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.The device was determined to meet functional and electrical performance specification requirements per rite testing.In addition, a visual inspection was performed.The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Upon conclusion of the investigation, the reported issue could not be verified.Should additional relevant additional information become available, a supplemental report will be submitted.
|