Model Number 5076-45
Device Problems Difficult to Remove (1528); Difficult to Advance (2920); Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/27/2015
Event Type  malfunction  
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: the full lead was returned, analysis was performed and no anomalies were found.Visual summary analysis of the lead indicated damage at implant and stretching.(b)(4).
 
Event Description
It was reported that during a difficult coronary sinus cannulation procedure using many guides and electrophysiology catheters, the right atrial (ra) lead became dislodged.There was difficulty advancing and removing the lead so it was explanted and replaced.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS
Type of Device
ELECTRODE, PACEMAKER, PERMANENT
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4972282
MDR Text Key23726444
Report Number2649622-2015-10235
Device Sequence Number1
Product Code DTB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P930039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Report Date 05/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/05/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/20/2017
Device Model Number5076-45
Device Catalogue Number5076-45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/27/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00078 YR
-
-