• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 6605632
Device Problems No Device Output (1435); High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/13/2015
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) replaced the rbc diluent dispenser due to micro bubbles, resolving the rbc recovery issues, and the instrument was verified as operational.The beckman coulter internal identifier for this report is (b)(4).
 
Event Description
The customer reported that red blood cell (rbc) results were not reproducing and were running high on a coulter lh 750 hematology analyzer; specifically, the rbc controls were out high and the rbcs on sample results were not matching other analyzers.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 750 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th avenue, 32-s08
miami, FL 33196-2031
3053802031
MDR Report Key4977171
MDR Text Key23563127
Report Number1061932-2015-01267
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Report Date 07/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/06/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6605632
Other Device ID NumberSOFTWARE VERSION: 2D2
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/13/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown