The subject device was returned to omsc for investigation.The evaluation.Based upon the evaluation of the subject device by omsc, the phenomenon duplicated.The subject device passed the leakage test.The service record was also reviewed and it was revealed that the subject device was assembled with the insertion unit fabricated in (b)(6) 2014.The manufacturing history was reviewed w/no irregularities related to this problem noted.The exact cause could not be conclusively determined at this time.If any add'l info becomes available at a later time, this report will be supplemented.
|
Olympus medical systems, corp.(omsc), was informed that after about one hour from the start of laparoscopic total gastrectomy, the image of the subject device disappeared.The subject device was replaced with a backup instrument that was same model w/the subject device.The procedure was completed and there was no patient harm.
|
This supplemental report is being submitted as additional information became available.Olympus disassembled the subject device and conducted further evaluation.There were no irregularities in the soldering of the video connector, and in the forming of inner cables.There were no buckling or cut in the cables.The outer appearance and the external diameter of the image unit were checked, with no irregularities found.There were no irregularities in the value of resistance and bias voltage.Adhesion area and its hardening condition, the cable unit and soldering of ccd were also checked, with no irregularities noted.Based upon the evaluation, as no irregularities were noted, damage in the ccd or the signal shaping circuit due to electrical stress such as static electricity could not be ruled out as a contributory factor of the event.
|