EDWARDS LIFESCIENCES PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD; CATHETER, OXIMETER, FIBEROPTIC Back to Search Results
Model Number 774HF75
Device Problem High Readings (2459)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/13/2015
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation; it was discarded at the hospital.Without return of the unit it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.It is not known if some procedural factors may have contributed to the event.No corrective actions will be taken at this time.A supplemental report will be forthcoming with the device history record results.
 
Event Description
It was reported that the catheter started out measuring properly and suddenly began measuring very high cardiac output values which were not related to the patient's clinical status.The monitor did not display any alarm.The clinician performed manual cold bolus's in order to evaluate cardiac output by thermodilution and was able to obtain values that were aligned with the patients status.The clinician did not take a picture and did not remember the values shown on the monitor.There was no more information available.There was no consequence for the patient as the user performed manual bolus.
 
Manufacturer Narrative
A review of the manufacturing records indicated that the product met specifications upon release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD
Type of Device
CATHETER, OXIMETER, FIBEROPTIC
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key4988008
MDR Text Key23862314
Report Number2015691-2015-01995
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Report Date 07/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/12/2016
Device Model Number774HF75
Device Lot Number59890181
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/13/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/14/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-