The referenced device was returned to olympus medical systems corp.(omsc) for evaluation.Omsc evaluated the device and found that the electrode of the md-631 seemed to be discolored due to high temperature.The high temperature was though to be due to insufficient application of the heat compound and/or inappropriate installation of the md-631.It would appear that the damage of the md-631 due to the high temperature caused the extinction of illumination, alternatively the spare lamp was lit and the error was displayed, and consequently the endoscopic image became dark.Based upon the finding of the evaluation, this report appears to be related to user handling.There were no further details provided.If significant additional information is received, this report will be supplemented.This report is being submitted as a medical device report in an abundance of caution.
|