Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 07/17/2015
Event Type  malfunction  
Manufacturer Narrative
Ge healthcare's investigation into the reported occurrence is still ongoing.A follow-up report will be issued when the investigation has been completed.
 
Event Description
During a planned visit, ge service representative noted the interlock system was not functional.There was no report of patient involvement.
 
Manufacturer Narrative
The unit was returned to the manufacturing site for investigation.The unit was inspected, and it was noted that the pivot pin and c-clip were missing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEC 6 PLUS
Type of Device
VAPORIZER
Manufacturer (Section D)
DATEX-OHMEDA
3030 ohmeda drive
madison WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI 53718
Manufacturer Contact
john szalinski
540 w. northwest highway
barrington, IL 60010-3076
8472774719
MDR Report Key4996440
MDR Text Key23847257
Report Number2112667-2015-00101
Device Sequence Number1
Product Code CAD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K000275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Followup
Report Date 07/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-