Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/17/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Ge healthcare's investigation into the reported occurrence is still ongoing.A follow-up report will be issued when the investigation has been completed.
|
|
Event Description
|
During a planned visit, ge service representative noted the interlock system was not functional.There was no report of patient involvement.
|
|
Manufacturer Narrative
|
The unit was returned to the manufacturing site for investigation.The unit was inspected, and it was noted that the pivot pin and c-clip were missing.
|
|