TELEFLEX MEDICAL HUDSON CIRCUIT, PEDIATRIC, DUAL LIMB ; BREATHING CIRCUIT Back to Search Results
Catalog Number 780-24
Device Problem Occlusion Within Device (1423)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/24/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product was not received at our facility.A device history record review could not be conducted since the lot number was not provided.No corrective action can be established since the sample is not available to perform an investigation and determine the source of defect reported.Customer complaint cannot be confirmed based only on the info provided, to perform an investigation and determine the source of defect reported it is necessary to evaluate the sample involved on this complaint.An attempt to duplicate the failure mode was made but at the time there is no inventory of the involved product code available at the facility nor is it being manufactured at the time.If device sample becomes available at a later date tis complaint will be re-opened.
 
Event Description
The customer alleges an occlusion issue due to crystallization of nebulized hypertonic saline within the neptune column on a ventilated pt.The occlusion caused a high pressure alarm on the vent.No pt harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CIRCUIT, PEDIATRIC, DUAL LIMB
Type of Device
BREATHING CIRCUIT
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo, tamaulipas 88275
MX   88275
Manufacturer Contact
margie burton, rn
3015 carrington mill blvd.
morrisville, NC 27560
9194334965
MDR Report Key4996520
MDR Text Key23834856
Report Number3004365956-2015-00230
Device Sequence Number1
Product Code OFP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Respiratory Therapist
Report Date 07/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number780-24
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/24/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-