ABBOTT POINT OF CARE I-STAT1 ANALYZER, IMMUNO READY Back to Search Results
Catalog Number 03P75-01
Device Problem Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/28/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
On (b)(6) 2015, abbott point of care (apoc) was contacted by a customer who reported that i-stat1 analyzer 369328 was yielding quality check code 44.The customer states that the rechargeable battery became hot to touch.There are no reports of smoke or burning smell.Product was replaced and the customer is returning both the analyzer and rechargeable battery for investigation.Apoc has determined that a component failure within the analyzer circuitry, may lead to the batteries becoming uncomfortably hot to touch in the area of the battery compartment when using a green non-fused battery carrier.However, the customer states that rechargeable batteries were being used.Therefore the analyzer is unlikely to become hot to touch.Based on the information available, there were no patient or user related injuries associated with this complaint.
 
Manufacturer Narrative
(b)(4).The investigation was completed on 10/21/2015.The failure was due to a defective tantalum capacitor.
 
Event Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT1 ANALYZER, IMMUNO READY
Type of Device
I-STAT1 ANALYZER
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
FLEXTRONICS MANUFACTURING (SINGAPORE)
pte ltd
1 kallang place
singapore 33921 1
SN   339211
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key4998070
MDR Text Key24704800
Report Number2245578-2015-00042
Device Sequence Number1
Product Code CGA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103195
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Followup
Report Date 10/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03P75-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/21/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown