• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY WALLSTENT (ILIAC, TIPS, VENOUS) ENDOPROSTHESIS WITH UNISTEP PLUS DELIVERY SYSTEM; STENT, CORONARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY WALLSTENT (ILIAC, TIPS, VENOUS) ENDOPROSTHESIS WITH UNISTEP PLUS DELIVERY SYSTEM; STENT, CORONARY Back to Search Results
Model Number UNK679
Device Problems Migration or Expulsion of Device (1395); Malposition of Device (2616)
Patient Problem Swelling (2091)
Event Date 07/17/2015
Event Type  Injury  
Manufacturer Narrative
Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.A review of the batch history, historical trending, and similar complaint trending review for the product family will be conducted.If there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Event Description
Same case as mdr id 2134265-2015-05788 and 2134265-2015-05789 and 2134265-2015-05250.It was reported that the patient experienced swelling of the leg.The target lesion was located in the iliac vein.A 18x40mm wallstent was deployed and post-dilated with good result.A second 18x40mm wallstent was deployed overlapping proximal to the previous stent.During post-dilation with two 18x40mm xxl esophageal balloon catheters, the physician noted that both balloons looked funny and inflated first at the "shoulders" way beyond the marks.During dilation the stent "jumped" about an inch and a half which made it to go into the inferior vena cava and no longer overlapping the previous stent.The procedure was concluded.Two days following the initial procedure the patient returned to the hospital due to swelling of the leg.Two balloon expandable stents were deployed inside the implanted wallstents.No further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLSTENT (ILIAC, TIPS, VENOUS) ENDOPROSTHESIS WITH UNISTEP PLUS DELIVERY SYSTEM
Type of Device
STENT, CORONARY
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5001439
MDR Text Key22899558
Report Number2134265-2015-05902
Device Sequence Number1
Product Code MAF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Report Date 07/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberUNK679
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/27/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-