• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number M004PM4500N40
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/08/2015
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).Device evaluated by manufacturer - the device has a kink at 13mm from the tip while in the neutral position (between ring 1 and 2).In addition the rings #1 and #2 has broken adhesive and fluids under them.The steering knob and the tension control knob functioned properly on both lock and unlock positions.The right and left curves are placed in the template shaded areas, the device passed the dimensional test.However, the device has a kink at the distal section.X ray analysis revealed that the center support is kinked.The distal section was dissected finding the center support kinked at 13 mm from the tip.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event Description
Reportable based on analysis completed on (b)(6) 2015.It was reported that the catheter was deformed.While performing a right atrial ablation, the distal tip of the intellatip mifi xp temperature ablation catheter became malformed.It appears that the steering mechanism of the catheter broke.The distal tip would not actively respond to straightening.The catheter was removed from the body multiple times in an effort to straighten the curve.Ultimately, the catheter was replaced with another intellatip mifi xp catheter and the case was completed.No patient complications were reported.However, analysis revealed broken adhesive between the electrodes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5002031
MDR Text Key22946934
Report Number2134265-2015-05265
Device Sequence Number1
Product Code OAD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P020025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Report Date 07/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/19/2016
Device Model NumberM004PM4500N40
Device Catalogue NumberPM4500N4
Device Lot Number16615976
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/21/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-