• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG OXYGENATOR, CARDIOPULMONARY BYPASS; DTZ

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG OXYGENATOR, CARDIOPULMONARY BYPASS; DTZ Back to Search Results
Model Number 70104.8759
Device Problem Material Deformation (2976)
Patient Problem Not Applicable (3189)
Event Date 08/10/2015
Event Type  malfunction  
Manufacturer Narrative
The product was investigated under complaint (b)(4) in the laboratory of the manufacturer.During visual inspection of the product a crack in the recirculation line connector was detected.This additional complaint was opened in order to cover the failure.No evidence was provided that this failure was noticed within the initial complaint report.The manufacturer's review of the quality control process indicated that a 100% functional inspection for leakage is conducted during manufacturing.The information obtained so far in this investigation would confirm that the device met its specification at the time of manufacturing and therefore all damages found on the product are due to excessive or inadequate external physical force that was exerted on the product after the release.The exact root-cause which led to the described failure could not be identified.This data will be handled through a designated maquet trending process.If a trend occurs, it will be escalated to quality assurance management for review and determination if further investigation is necessary.Due to this no further action will be completed at this time.
 
Event Description
During investigation of a complaint, an additional malfunction was found in the laboratory of the manufacturer.During visual inspection of the product a damaged blood outlet connector was detected.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXYGENATOR, CARDIOPULMONARY BYPASS
Type of Device
DTZ
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer Contact
michael campbell
kehler strabe 31
rastatt 76437
GM   76437
2229321132
MDR Report Key5008671
MDR Text Key23519617
Report Number8010762-2015-00887
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K082117
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Not Applicable
Report Date 08/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/01/2016
Device Model Number70104.8759
Device Catalogue NumberHMO 70000-J
Device Lot Number70101041
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/17/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received08/10/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-