(b)(4).The carefusion failure analysis engineer examined the 3100b and found that the bias flow could not be adjusted to 30 lpm and therefore could not be set to the ventilator performance checks settings.Determined the pr7 was out of calibration.Calibrated pr7 to a maximum bias flow reading of 60+/-1 lpm per test and servicing of 3100b procedure.Noted that the pr7 regulator made a squeaking sound as it was being adjusted.Operated unit at ventilator performance checks settings for 4 hours with no problems.Finding/root-cause: could not duplicate, the bias flow was fluctuating, complaint allegation.
|