DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.010 |
Device Problem
Difficult or Delayed Activation (2577)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/14/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was confirmed.An assessment was performed and it was observed that the device did not function.It was further noted that the electric motor and electronic control unit were damaged/not functional and the upper trigger was sticking.The assignable root cause was determined to be due to component wear from normal use over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported that the small battery drive device was running very slow.The reporter stated that the device was used for training purposes only.The event was not related to surgery.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|