Catalog Number ER320 |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/22/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Information anticipated, but unavailable at this time.When additional information is received and/or the device analysis has been completed, a supplemental medwatch will be sent.Additional information was requested and the following was obtained: what is meant by clips not seating properly? customer said they would not close all the way.Did clips not feed into jaws or partially feed? n/a.Did clips feed sideways? n/a.Did clips slow feed? n/a.Did multiple clips feed? customer said a few fires multiple clips were fed.
|
|
Event Description
|
It was reported that during an unknown procedure, the customer said that the clips were not seating properly.An er420 device was used to complete the procedure.There were no adverse consequences for the patient.
|
|
Manufacturer Narrative
|
(b)(4).Should the information be provided later, a supplemental medwatch will be sent.Batch # m90v8l.The er320 device was returned for analysis and upon inspection the jaws were found to be in a yielded condition.In an attempt to replicate the reported incident the device was functionally evaluated.Upon firing of the device, the remaining clips were ejected due to the condition of the jaws; finally the instrument locked out as intended.Possible causes for the found condition of the yielded jaws may be if the device is closed over an existing hard object or clip placing stress on the jaws causing them to distort or yield and not return to their original dimensions/position or excessive application of torque to the jaws when positioning the device on a vessel.The batch record was reviewed and no anomalies were noted during the manufacturing process.
|
|
Search Alerts/Recalls
|