ETHICON ENDO-SURGERY, LLC. PDS CORD; CLIP, IMPLANTABLE Back to Search Results
Catalog Number ER320
Device Problem Fitting Problem (2183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/22/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Information anticipated, but unavailable at this time.When additional information is received and/or the device analysis has been completed, a supplemental medwatch will be sent.Additional information was requested and the following was obtained: what is meant by clips not seating properly? customer said they would not close all the way.Did clips not feed into jaws or partially feed? n/a.Did clips feed sideways? n/a.Did clips slow feed? n/a.Did multiple clips feed? customer said a few fires multiple clips were fed.
 
Event Description
It was reported that during an unknown procedure, the customer said that the clips were not seating properly.An er420 device was used to complete the procedure.There were no adverse consequences for the patient.
 
Manufacturer Narrative
(b)(4).Should the information be provided later, a supplemental medwatch will be sent.Batch # m90v8l.The er320 device was returned for analysis and upon inspection the jaws were found to be in a yielded condition.In an attempt to replicate the reported incident the device was functionally evaluated.Upon firing of the device, the remaining clips were ejected due to the condition of the jaws; finally the instrument locked out as intended.Possible causes for the found condition of the yielded jaws may be if the device is closed over an existing hard object or clip placing stress on the jaws causing them to distort or yield and not return to their original dimensions/position or excessive application of torque to the jaws when positioning the device on a vessel.The batch record was reviewed and no anomalies were noted during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDS CORD
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 00969
Manufacturer Contact
milton garrett
4545 creek road ml 120a
cincinnati, OH 45242
5133378865
MDR Report Key5026047
MDR Text Key24780502
Report Number3005075853-2015-05280
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Report Date 07/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/27/2020
Device Catalogue NumberER320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/26/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2015
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-