It was reported that during a shoulder procedure using a super turbovac 90 icw wand, the outer ring detached from the wand tip while inside the surgical site.The detached piece was removed and the procedure was completed successfully, however, the event caused a 30 minute surgical delay.There have been no reported patient complications.
|
The complaint could not be verified nor could a definitive root cause be determined with confidence as the device was not returned for analysis.There was no information available to indicate what the total activation time was, set points used or whether a cannula or any other apparatus¿ were used during the procedure.Factors not associated with the manufacture or design of the device which could result in damage to the tip could be that the device came into abrupt contact with a hard (metallic) object, improper removal from an apparatus, and use at a higher than recommended set point or excessive force applied to tip can also lead to unintended.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
|