Model Number MODEL 250
Device Problems Image Display Error/Artifact (1304); Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/24/2015
Event Type  malfunction  
Event Description
It was reported that the physician's tablet would not power on.Once the tablet finally did power on, the screen was noted to be full of lines.The complaint tablet was received on (b)(6) 2015.The physician was provided a new programming tablet.Device analysis is underway but has not been completed to date.
 
Event Description
Product analysis was performed which showed the cause of the reported power on issues was associated with an upside down power button.Once the power button was installed correctly, no further anomalies associated with the tablet were identified.The dhr was reviewed and it was noted the device passed all functionality tests and inspection.
 
Manufacturer Narrative
The udi # was inadvertently left of the original mfr.Report; (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5027315
MDR Text Key24796923
Report Number1644487-2015-05603
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Report Date 07/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2015
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received10/06/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/21/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-