• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR Back to Search Results
Model Number LC100
Device Problems Self-Activation or Keying (1557); Smoking (1585)
Patient Problem No Code Available (3191)
Event Date 08/07/2015
Event Type  Injury  
Manufacturer Narrative
The device has not been made available for evaluation at this time.Should further information or the device become available for evaluation, a follow-up report will then be issued.
 
Event Description
Customer alleges chair was working without pressing the hand control then smoke started coming out of motor.The house was filled with smoke and fire department came and got customer out of chair.
 
Manufacturer Narrative
The device was returned and evaluated.A broken wire was found.The device has been in the field for two (2) years and it is unknown how the wire broke.
 
Event Description
Customer alleges chair was working without pressing the hand control then smoke started coming out of motor.The house was filled with smoke and fire department came and got customer out of chair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Type of Device
ELECTRIC POSITIONING CHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
182 susquehanna ave
n/a
exeter PA 18643
Manufacturer (Section G)
N/A
n/a
n/a
n/a
Manufacturer Contact
kelly livingston
182 susquehanna ave
n/a
exeter, PA 18643
5706555574
MDR Report Key5029377
MDR Text Key24010932
Report Number2530130-2015-00067
Device Sequence Number1
Product Code INO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091578
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Report Date 09/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberLC100
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received08/14/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-