According to the investigation, the cause of the table running on its own was damage to the keypad in the remote.This issue is dealt with in the ongoing fsca ((b)(4)).The keypad used was prior to the design change.Trumpf medical has issued safety instructions to the users in the field safety notice and those should be followed until trumpf medical is able to do the final exchange.The operation table and the remote control were checked by service and repaired.Anticipated completion of the fsca is july 2016.
|