Engineering analysis: the sample was removed from the bio-hazard bag and inspected to determine the cause of the complaint.On initial inspection the stent was no longer on the balloon and was shifted proximally up the catheter shaft.The balloon was still in the folded position and showed no signs of contrast fluid in the proximal or distal balloon cones.The stent and catheter showed no signs of bodily fluids, not even at the distal tip that would have entered the sheath prior to the stent making contact with the valve of the introducer sheath.The balloon surface was evaluated to determine if the stent was crimped properly during manufacturing, when the stent is crimped on to the folded balloon the stent frame leaves impressions on the surface of the balloon.The impressions indicate if the stent was properly crimped.The impressions were clearly visible indicating that the stent was properly crimped by manufacturing.The diameter of the stent was measured.The diameter was 2.18mm.This diameter is in line with the 59 samples measured during the product lot qualification testing.It is only slightly larger due to traveling over the proximal balloon cone when dislodged.The sheath used in the case was not returned.As the details indicate, this device was used in conjunction with an aaa endograft.It is possible that the sheath was damaged prior to advancing the icast covered stent into the valve of the sheath.This cannot be confirmed without the introducer sheath.The icast instructions for use specify the following, "do not force passage or withdrawal of the guidewire or delivery system if resistance is encountered." engineering summary: a full review of the catheter lot history records for the device in question was performed.The records indicate that this lot of catheters passed atriums final lot qualification testing.This inspection requires that the catheter lot must pass the following: ability of the stent and delivery system to be passed through the labeled introducer sheath.Ability to deploy the stent at nominal pressure (8atm).Ability to withdraw the deflated balloon catheter back through the labeled introducer sheath ability of the delivery system to withstand 10 inflate/deflate cycles at the rated burst pressure (12atm) without leaks or failures.Balloon burst testing.The balloon must burst over the rated burst pressure specified on the label (12atm) manifold to shaft tensile testing.Samples when tensile tested must not break or separate below 3.37lbs result: all (b)(4) quality inspection samples passed this final inspection without any non-conformances noted.We have not been able to determine any fault due to manufacturing.
|