Model Number 95257
Device Problems Break (1069); Scratched Material (3020)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 08/28/2015.One scd controller compression system was returned for a damaged power cord.Upon triage at a local service center it was noticed there were exposed copper wires on the power cord.Initial inspection of the power cord showed it failed to meet operational specifications because it was externally damaged, exposing the internal copper wires which presented an electrical shock hazard, confirming there was an issue with the power cord.The root cause of the power cord failure can be attributed to rough handling.Power cords periodically require replacement due to age, usage and user damage.The service manual instructs the user to periodically inspect the power cords resistance to ensure its electrical safety.The power cord was replaced to correct the problem.The scd controller was fully tested and passed all operational specifications.The scd controller was manufactured in 2007.A review of the device history records shows this device was released meeting all manufacturing specifications.
 
Event Description
The customer stated the controller had a damaged power cord.No patient involvement.The extent of the issue with the power cord was unknown to covidien at the time the customer reported the issue.The scd controller was sent to a covidien service center and on upon triage the service technician noted that there was damage to the power cord that consisted of exposed copper wires.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCD EXPRESS COMP SYSTEM JAPAN
Type of Device
SCD CONTROLLER
Manufacturer (Section D)
COVIDIEN
2824 airwest boulevard
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN
2824 airwest boulevard
plainfield IN 46168
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key5040465
MDR Text Key25219566
Report Number3008361498-2015-00008
Device Sequence Number1
Product Code JOW
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Report Date 08/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number95257
Device Catalogue Number95257
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/07/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-