DEPUY SYNTHES POWER TOOLS UNIVERSAL BATTERY CHARGER II; BATTERY, REPLACEMENT, RECHARGEABLE
|
Back to Search Results |
|
Catalog Number 05.001.204 |
Device Problems
Failure to Charge (1085); Corroded (1131); Crack (1135)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/18/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned for service, however did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and observed that the device would not charge the battery devices, there was corrosion on the contacts and the feet on the bottom of the device were cracked.Therefore, the reported condition was duplicated and confirmed.The assignable root cause was determined to be due to wear on components from use over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that during service and repair pre-testing, it was discovered that the universal battery charger device would not charge the battery devices, the feet on the bottom were cracked and there were corrosions on the contacts.The event was not related to surgery.There was no patient involvement.There were no injuries or medical intervention associated with this event.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|