Catalog Number 5C8310R
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Electric Shock (2554)
Event Date 08/12/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received for evaluation.The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Internal and external inspection was performed and no issues were noted.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.The device was determined to meet functional and electrical performance specification requirements per rite testing.The reported problem of electric shock was neither verified nor duplicated during evaluation and the cause was undetermined.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a caregiver experienced an electric shock from the homechoice device.This occurred while the caregiver was moving the homechoice for the patient.The patient was not connected for therapy and the homechoice was unplugged.The technical service representative swapped the device.The patient would perform therapy with manual supplies.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5042403
MDR Text Key24611499
Report Number1416980-2015-34298
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Report Date 08/31/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C8310R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2015
Date Manufacturer Received08/12/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age51 YR
-
-