(b)(4).The device was received for evaluation.The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Internal and external inspection was performed and no issues were noted.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.The device was determined to meet functional and electrical performance specification requirements per rite testing.The reported problem of electric shock was neither verified nor duplicated during evaluation and the cause was undetermined.Should additional relevant information become available, a supplemental report will be submitted.
|