BECKMAN COULTER ACCESS 2 IMMUNOASSAY SYSTEM; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
|
Back to Search Results |
|
Catalog Number 81600N |
Device Problem
Calibration Problem (2890)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/05/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer did not provide patient demographics such as age, date of birth, sex or weight.Beckman coulter (bec) customer technical support (cts) contacted the customer's onsite bio med service engineer.The engineer stated after replacement of the main pipettor/transducer the engineer ran water as negative samples and all access accutni+3 results were recovering as expected.The cause of the reported event is due to a malfunction of the main pipettor/transducer.
|
|
Event Description
|
The customer stated they noted indeterminate (ind) flagged troponin i (access accutni+3) results for four (4) patients on the laboratory's access 2 immunoassay system ( serial number (b)(4)).The customer repeat tested one (1) sample two (2) times on the same access 2 immunoassay system and obtained two (2) additional ind-flagged access accutni+3 results the purpose of the ind result flag is to alert the customer that a sample reading is outside the range of the current assay calibration curve and cannot be distinguished from a system or operator error and indicates that a final result cannot be calculated.It is provided to prevent reporting of potentially erroneous results, which it did in this case as intended.There was no report of any change or impact to patient care or treatment.Level one (1) quality control (qc) was recovering outside the customer's established limits.A precision run using low level qc material failed to meet performance specifications.Sample information, such as collection tube type, centrifugation time and speed, were not provided.The customer did not note any issues with sample integrity.
|
|
Search Alerts/Recalls
|
|
|