• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES BRANDYWINE TI S-HOOK-RIGHT; PROSTHESIS, RIB REPLACEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES BRANDYWINE TI S-HOOK-RIGHT; PROSTHESIS, RIB REPLACEMENT Back to Search Results
Catalog Number 497.258
Device Problems Break (1069); Difficult to Remove (1528)
Patient Problems Failure of Implant (1924); Sedation (2368)
Event Date 08/12/2015
Event Type  Injury  
Manufacturer Narrative
Exact weight reported as (b)(6).The investigation could not be completed; no conclusion could be drawn, as no product was received.Device history record review: manufacturing date: 17april2012.Part 497.258, lot 6879222 did not contain any non-conformance reports or any anomalies.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that while resetting/replacing a left hook during a scheduled revision, it was discovered that the right hook implant was broken below the titanium parallel connector and plowing through the bone.The hook was difficult to remove as device was embedded in the patient's bone.Surgery time was prolonged by thirty to forty-five (30-45) minutes.Patient status is reported as okay.The original date of implantation is unknown.This is report 1 of 2 for com-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI S-HOOK-RIGHT
Type of Device
PROSTHESIS, RIB REPLACEMENT
Manufacturer (Section D)
SYNTHES BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer (Section G)
SYNTHES BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5049646
MDR Text Key24879859
Report Number2530088-2015-10576
Device Sequence Number1
Product Code MDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK142587
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Report Date 08/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/02/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number497.258
Device Lot Number6879222
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/24/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/17/2012
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age17 YR
Patient Weight41
-
-