Catalog Number A000823000100
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that a gate drill separated; the separated piece was retrieved and no injury resulted.
 
Manufacturer Narrative
Multiple unsuccessful attempts were made to obtain the device for evaluation.A dhr review was conducted with no discrepancies noted.Root causes are not identified.We will track this kind of event and monitor the trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GATES GLIDDEN DRILL
Type of Device
DRILL, DENTAL, INTRAORAL
Manufacturer (Section D)
DENTSPLY MAILLEFER
chemin du verger 3
ballaigues, 1338
SZ  1338
Manufacturer (Section G)
DENTSPLY MAILLEFER
chemin du verger 3
ballaigues, 1338
SZ   1338
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key5051164
MDR Text Key25498607
Report Number8031010-2015-00038
Device Sequence Number1
Product Code DZA
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 08/04/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/02/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA000823000100
Device Lot Number1086504
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/07/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-