Catalog Number 25382
Device Problem Detachment Of Device Component (1104)
Patient Problem Failure of Implant (1924)
Event Date 02/21/2015
Event Type  Injury  
Manufacturer Narrative
Because this event resulted in medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, it must be presumed that the malfunction would be likely to cause or contribute to a serious injury should it recur.As such, this event meets the definition of a reportable event per 21 cfr part 803.Evaluation found significant corrosion on the laser marking, the driver shaft is slightly bent and the contra-angle-lock was damaged.
 
Event Description
In this event it was reported that an ev implant driver 4.8 (short) was bent and possibly caused an implant to fall off the implant driver.Therefore, the surgery could not be completed.The patient returned one week later and the surgery was successfully completed with a new implant driver.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANT DRIVER EV 4.8 SHORT
Type of Device
ACCESSORIES, IMPLANT, DENTAL, ENDOSSEOUS
Manufacturer (Section D)
DENTSPLY IMPLANTS (A DIVISION OF DENTSPLY IH AB)
aminogatan 1
molndal, vastra gotalands lan [se-14] S-431 21
SW  S-431 21
Manufacturer (Section G)
DENTSPLY IMPLANTS (A DIVISION OF DENTSPLY IH AB)
aminogatan 1
molndal, vastra gotalands lan [se-14] S-431 21
SW   S-431 21
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key5051264
MDR Text Key24879878
Report Number9612468-2015-00012
Device Sequence Number1
Product Code NDP
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dentist
Report Date 06/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/02/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number25382
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/08/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age57 YR
-
-