• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number 3200
Device Problems Device Alarm System (1012); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/13/2015
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the external buzzer blown.Per the hill-rom user manual, warning: the bed exit alarm system is not intended as a substitute for good nursing practices.The bed exit alarm system must be used in conjunction with a sound risk assessment and protocol.Per the hill-rom user manual, if the bed exit alarm does not arm and all three mode indicators are flashing, remove the patient and zero the bed exit system.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2014 and 2015.It is unknown if the facility performed any other preventative maintenance on this bed.Hill-rom technician replaced the buzzer to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the brake not set alarm was inoperable.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
john cummings
1069 state route 46 east
batesville, IN 47006
8129312869
MDR Report Key5052778
MDR Text Key25499611
Report Number1824206-2015-00862
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Report Date 08/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received08/13/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/03/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-