CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problem Inaccurate Delivery (2339)
Patient Problem No Patient Involvement (2645)
Event Date 08/06/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The carefusion field service engineer evaluated the ventilator and discovered map adjustment knob was turned past stop post (out of adjustment).Repaired map knob and calibrated ddi board and power module.Performed testing/checkout.Verified alarms for all parameters.(b)(4).
 
Event Description
The customer reported that when the map gets to about 9cm that the driver seems to not run smoothly, seems to "stutter".It was found upon calibrations.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREFUSION
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
jill rittorno
75 n. fairway drive
vernon hills, IL 60061
8473628056
MDR Report Key5053383
MDR Text Key25427945
Report Number2021710-2015-01559
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Health Professional
Report Date 08/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/06/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-