• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROTEGE IPG, 16-CHANNEL RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROTEGE IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3789
Device Problems Charging Problem (2892); Unintended Movement (3026)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/14/2015
Event Type  Injury  
Manufacturer Narrative
Patient code: (b)(4).Device code: (b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.No udi available.This medical device report was initially submitted on (b)(6) 2015 and is being resubmitted per request of regulatory agency.
 
Event Description
It was reported the patient was experiencing difficulty in recharging her ipg.The patient had to press hard against the ipg with the charging antenna to ensure there was a connection.It was determined the ipg was slanted in the pocket with the header protruding.Follow up information identified the ipg pocket was revised on 7/15/2015 which resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGE IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
tanja hoffman
6901 preston road
plano, TX 75024
MDR Report Key5056861
MDR Text Key214723067
Report Number1627487-2015-26355
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/04/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2017
Device Model Number3789
Device Lot Number5006847
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/14/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/10/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-