Model Number N/A
Device Problems Fracture (1260); Metal Shedding Debris (1804)
Patient Problems Pain (1994); Complaint, Ill-Defined (2331)
Event Date 08/13/2015
Event Type  Injury  
Manufacturer Narrative
The product identification necessary to review manufacturing history was not provided.Current information is insufficient to permit a conclusion as to the cause of the event.
 
Event Description
It was reported that the patient had an initial shoulder procedure on (b)(6) 2010.Subsequently, the patient was closing the trunk of a car and felt pain.The patient was revised due to a fractured trunnion on (b)(6) 2015 where it was discovered the patient had metallosis and a pseudocapsule.
 
Manufacturer Narrative
This follow-up report is being filed to relay additional information, which was unknown at the time of the initial medwatch.Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.There are warnings in the package insert that state that this type of event can occur: under possible adverse effects, number 1 states, ¿material sensitivity reactions.¿ number 10 states, "fatigue fracture of component can occur as a result of loss of fixation, strenuous activity, malalignment, trauma, non-union, or excessive weight." number 13 states, "intraoperative or postoperative bone fracture and/or postoperative pain.".
 
Manufacturer Narrative
This follow-up report is being filed to correct information.Upon reassessment of the reported event, it was determined the event is attributed to pn: 155340 ln: 310600.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP RVS HMRL TI TRAY 44MM
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
megan haas
56 e. bell drive
warsaw, IN 46582
5743726700
MDR Report Key5057185
MDR Text Key25037605
Report Number0001825034-2015-03866
Device Sequence Number1
Product Code PAO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK080642
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Physician
Type of Report Followup
Report Date 10/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/04/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date05/31/2020
Device Model NumberN/A
Device Catalogue Number115340
Device Lot Number310600
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/06/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/03/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age80 YR
-
-