• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Failure to Interrogate (1332); Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/01/2015
Event Type  malfunction  
Event Description
It was reported that a medical professional could not interrogate generators when using her programming system.The programming wand was used with different tablets, but the problem persisted.Further information was later received indicating that further troubleshooting determined that the usb cable was suspected to be defective.The return of the suspect usb cable is expected but it has not been received to date.Review of manufacturing records confirmed that the tablet passed all functional tests prior to distribution.
 
Event Description
The suspected usb cable was returned to the manufacturer on 09/10/2015.Analysis performed on the returned usb cable, confirmed broken lead at db9 interface of db9 to usb serial adapter communication cable.The cause for the reported allegation is associated with a disconnected wire connection in the returned serial cable, which appears to be cable stress-related.Once the wire was soldered onto the serial cable pcb, no further anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5063258
MDR Text Key25921741
Report Number1644487-2015-05716
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2015
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received09/10/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-