• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500; RP 500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500; RP 500 Back to Search Results
Catalog Number 10492730
Device Problem Computer Operating System Problem (2898)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/12/2015
Event Type  No Answer Provided  
Manufacturer Narrative
Siemens is in process of investigating this event.The cause for the event is unknown at this moment.
 
Event Description
Customer reported analyzer populated operator id on the patient id field on the demographic screen while running blood samples.Customer indicated that on (b)(6) 2015; they were running blood samples after installing new measurement cartridge and noticed that her operator id ((b)(6)) populated on the patient id field before scanning or typing anything on the patient id field.There was no report of serious injury due to this event.
 
Manufacturer Narrative
Siemens software engineering team reviewed the complaint and was unable to reproduce the event.Customer has not indicated any additional occurrences.There also have not been any previous documented occurrences of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key5068858
MDR Text Key26114663
Report Number1217157-2015-00131
Device Sequence Number1
Product Code KHP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Respiratory Therapist
Report Date 10/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Catalogue Number10492730
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/02/2015
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
Patient Age36 YR
-
-