Procedure: laparoscopic sigmoid colectomy.According to the reporter: clearify visualization system ((b)(4)) wand tip broke off during a laparoscopic sigmoid colectomy.The tip was not found.There was no unanticipated tissue loss.The incision was not extended by more than one inch.There was no unanticipated blood loss of 500cc or more.Surgery time was not delayed by more than 30 minutes.Delay in surgery? no (i.E.An extension of the hospital stay, infection, etc.?) 9.What is the current status of the patient? pt.Expired not related to device 10.Can you confirm that the device will be returned for evaluation? unknown where the device goes after sent to medsun 11.Was the device reprocessed or re-sterilized prior to use? no 12.Did you experience any issue with the endopath xcel trocar or was just used in the same procedure? no if so, please describe any problems with the trocar.Additional information requested via email from (b)(6) the medwatch form stated that the lot number was p5a0380x but the response below states that the lot number is j0814hx.I will update the file accordingly with the appropriate lot number and that should be the last piece of information i need.Account confirmed that lot number is actually j0814hx.
|