Driveline was inspected during service and remains implanted.The controller evaluated during service, is not expected to be returned to manufacturer.Log analysis - for general information purposes- con099201 log file analysis review date: august 13, 2015.Data through: august 13, 2015.Normal power consumption.1 electrical fault alarm was logged on august 12, 2015.The heartware vad is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings, including transportation.The instructions for use (ifu) outlines required surgical steps to prevent driveline contamination during tunneling.It additionally warns that failure to follow instructions on protecting the driveline connector or improper use of the driveline cap could result in contamination or damage to the connector and electrical fault alarms could occur.The ifu and patient manual also include a reference guide for alarms including potential causes and actions to take.Heartware will submit a supplemental report when new facts arises which materially alters information submitted in a previous mdr report.This is one of two reports (3007042319-2015-02225 and 3007042319-2015-02226) submitted for devices related to the same event.
|
The driveline cable was not returned for evaluation.Review of the manufacturing documentation confirmed that the associated device met all requirements for release.The reported event was confirmed via review of the controller log files, which revealed 1 electrical fault.The electrical fault alarm may be indicative of compromised inner lumen wires, or an overcurrent condition due to an electrical short in the driveline or driveline connector as a result of the damage sustained to the driveline.Furthermore, on-site inspection of the driveline cable revealed a deep cut through the outer sheath and partial cut in the inner lumen which aligns with the log file analysis.The confirmed malfunction is related to the reported event.The most likely root cause of the reported electrical fault alarm can be attributed to a severed driveline.The most likely root cause of the driveline sheath damage is exposure to uv light.Heartware has opened an internal investigation to evaluate this type of issue.The most likely root cause of the reported electrical fault alarm can be attributed to a severed driveline.The most likely root cause of the driveline sheath damage is exposure to uv light.Heartware will submit a supplemental report when new facts arises which materially alters information submitted in a previous mdr report.This is one of two reports (3007042319-2015-02225 and 3007042319-2015-02226) submitted for devices related to the same event.
|