PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. INGENUITY CT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
|
Back to Search Results |
|
Catalog Number III |
Device Problems
Device Alarm System (1012); Leak/Splash (1354)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.(b)(4).
|
|
Event Description
|
In this case, the customer reported that the charger had failed on the uninterruptable power supply (ups) causing an alarm to sound.A philips field service engineer (fse) confirmed there was no harm to a patient, operator, or bystander.The fse reported that there was battery acid leakage from the ups batteries.The fse replaced the ups to resolve the issue.
|
|
Manufacturer Narrative
|
On (b)(6) 2015 the customer reported that the charger had failed on the console uninterruptible power supply (ups) causing an alarm to sound.A philips field service engineer (fse) confirmed there was no harm to a patient, operator, or bystander.The philips fse stated that there were no log files pulled as a result of this event as well as the defective console ups was scrapped after this event occurred.The philips fse stated that the failed console ups was the original unit that was installed with the ct system and it was 3 years old at the time of this event.The console ups was replaced and the system is operating as intended.Ct engineering determined this issue to be an acceptable risk.-the system shall only utilize batteries approved by a recognized electrical safety organization (i.E.Ul, csa).-ups owner¿s manual states no user serviceable parts and requires no routine maintenance.-full system ups has temperature sensor monitoring the battery cabinet temperature.-ups comes with status notification (beep alarm for fault condition).-service and user documentation shall identify appropriate personal protective equipment (ppe) and neutralizing agents (ammonia or baking soda).-site planning document recommends to not allow ups closer than 1.5m (60 inches) from patient couch unless it is certified as a medical grade device and/or approved with the ct system.-site planning document recommends that a ups battery cabinet should not be exposed to prolonged periods of temperature above 25 c (77 f).The probable cause based upon the information available is due to defective batteries in the console ups was due to the batteries being 3 years old and needed to be replaced.
|
|
Search Alerts/Recalls
|
|
|