INVACARE FLORIDA OPERATIONS PERFECTO2 OXYGEN CONCENTRATOR WITH SENSO2 9153646963; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number IRC5PO2
Device Problem Cut In Material (2454)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Should additional information become available, a supplemental record will be filed.
 
Event Description
Dealer states unit has a cut power cord.
 
Manufacturer Narrative
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that there was no malfunction of the power cord indicated, rather it was replaced at the request of the dealer, so the original complaint issue could not be confirmed.However, additional malfunctions were found with the device, including a defective pilot valve and stuck 4-way valve which were causing the unit to alarm.
 
Event Description
Dealer states unit has a cut power cord.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFECTO2 OXYGEN CONCENTRATOR WITH SENSO2 9153646963
Type of Device
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer (Section G)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5090513
MDR Text Key26556818
Report Number1031452-2015-16055
Device Sequence Number1
Product Code CAW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Repair
Type of Report Followup
Report Date 10/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberIRC5PO2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received10/22/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-