The reported incident of the pin breakage could not be confirmed, since the device was not returned for evaluation.Based on the investigation, the root cause was attributed to a r&d/design related issue.R&d maintenance took over this failure mode within the framework of a potential recurring situation identification board and reported the following: "the potential to improve the insertion behaviour shall be investigated".(b)(4).Tolerance field of the tip is to be reduced.(b)(4) has been opened and will address this failure mode.A review of the device history for the reported lot did not indicate any abnormalities.A review of the labeling did not indicate any abnormalities.No indications of material or manufacturing related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|