Model Number ADVIA CENTAUR XP |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/27/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
A siemens customer service engineer (cse) specialist was dispatched to the customer site.After evaluating the instrument, the cse checked the probe alignment and found that the sample probe connector was loose.The cse replaced the sample probe and performed sample probe pressure test, which was acceptable.The cse also ran quality controls and patient samples.The cause of the discordant ferritin, folate and vitamin b12 results on multiple samples is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|
|
Event Description
|
Discordant ferritin (ferr), folate (fol) and vitamin b12 (vb12) results were obtained on multiple patient samples on an advia centaur xp instrument.The discordant results were not reported to the physician(s).The samples were repeated on the same instrument, resulting different from the initial results.The repeat results were reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant ferritin, folate and vitamin b12 results.
|
|
Manufacturer Narrative
|
The initial mdr 2432235-2015-00411 was filed on september 23, 2015.Additional information (09/15/2015): upon a follow-up visit, a siemens customer service engineer (cse) specialist found microbubbles at the wash manifold line and suspected micro leak at the connectors.The cse replaced the reagent probe 1, the reagent probe 2 and the wash displacement line at the manifold.The cse ran calibrations and quality controls, which were acceptable.A siemens headquarters support center (hsc) specialist reviewed the service report and indicated that the cause of the discordant ferritin, folate and vitamin b12 results on multiple samples was related to a malfunction of the manifold connectors.The instrument is performing according to specifications.No further evaluation of the device is required.
|
|
Search Alerts/Recalls
|