STRYKER MEDICAL LONDON DBA CHG HOSPITAL BEDS SPIRIT SELECT; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Catalog Number 5700000000
Device Problem Device Sensing Problem (2917)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/27/2015
Event Type  malfunction  
Event Description
It was reported there was an issue with the obstruction sensor.No patient was affected and no adverse consequences were reported.
 
Manufacturer Narrative
Follow-up submitted as this complaint was determined not to be reportable.An obstruction sensor malfunction could result in the lift being stuck, however, it is not likely to harm the patient.After the patient was discharged, however, it was found that the bed had been soaked during cleaning and once the bed dried it worked correctly.There was no defect found and the unit was put back into service.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.This issue is not likely to cause or contribute to serious injury or death if it was to recur.
 
Event Description
It was reported there was an issue with the obstruction sensor.No patient was affected and no adverse consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIRIT SELECT
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
STRYKER MEDICAL LONDON DBA CHG HOSPITAL BEDS
1020 adelaide st. s.
N6E 1 R6
CA  N6E 1R6
Manufacturer (Section G)
STRYKER MEDICAL LONDON DBA CHG HOSPITAL BEDS
1020 adelaide st. s.
N6E 1 R6
CA   N6E 1R6
Manufacturer Contact
rita moffitt
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key5095694
MDR Text Key26696272
Report Number3006433555-2015-00393
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Followup
Report Date 08/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number5700000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received08/27/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-