• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVACARE REHABILITATION EQUIPMENT CO. POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI; WHEELCHAIR, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE REHABILITATION EQUIPMENT CO. POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI; WHEELCHAIR, POWERED Back to Search Results
Model Number M41SRR
Device Problems Detachment Of Device Component (1104); Mechanics Altered (2984)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Should additional information become available a supplemental record will be filed.
 
Event Description
Right motor disengaging intermittently and states while on a ramp, the right motor completely locked up and the end user was almost thrown off the chair.
 
Manufacturer Narrative
Per the documentation on the irs or return fields in oracle, the failure could not be duplicated, motor operated properly during bench test.Wiggled wires and brake handle with no failure.
 
Event Description
Right motor disengaging intermittently and states while on a ramp, the right motor completely locked up and the end user was almost thrown off the chair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI
Type of Device
WHEELCHAIR, POWERED
Manufacturer (Section D)
INVACARE REHABILITATION EQUIPMENT CO.
no.5 weixi road, sip
suzhou jiangsu 21512 1
CH  215121
Manufacturer (Section G)
INVACARE REHABILITATION EQUIPMENT CO.
no.5 weixi road, sip
suzhou jiangsu 21512 1
CH   215121
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5097233
MDR Text Key26767231
Report Number3008262382-2015-01732
Device Sequence Number1
Product Code ITI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Followup
Report Date 09/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberM41SRR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received09/17/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-