This event has been reassessed as not mdr reportable.It was originally reported that the cool flow pump failed to activate high flow irrigation during ablation.However, additional information was received that indicates that the stockert generator sensed that the pump did not spool up and did not allow for radiofrequency ablation energy to be delivered during that time.As such, the device functioned as intended because if there is no flow the generator cannot deliver radiofrequency.The potential that this could cause or contribute to a death or serious injury, or other significant adverse event, is remote.Due to the (b)(4) 2015 fda maintenance where the 3500a codes were updated, the 3500a codes will be added until the biosense webster system is also updated.Therefore the following codes apply: (b)(4).Manufacturer's ref.No: (b)(4).It was reported that the pump intermittently failed to start when radiofrequency was switched on.In addition, the bubble alarm seemed to go off constantly as well as the pump door seemed to be very stiff.The device was evaluated.Bu error was confirmed.Door was difficult to open.Rf error was not duplicated.Bubble sensor and door assembly were replaced and the issue was resolved.The device was also subjected to preventative maintenance, safety and functional testing and all tests passed.
|