Because this issue could possibly lead to file separation, which could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, it must be presumed that the issue would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device arrived in very bad conditions.The problem was due to the foot pedal; it was reworked by somebody, it is clear that on the switch there's a long scratch and a kind of "depression" in the plastic shell that create a shortcut inside.In fact removing the plastic shell the pedal returned to work normally.
|