• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problem Device Difficult to Program or Calibrate (1496)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/17/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Problem: the carefusion field service engineer evaluated the ventilator and could not find any issue.Evaluated using the settings provided by the staff and found no issue.Ventilator performance check was performed and also checked inside the unit for any mucus or any water in the tubes, nothing found.Patient circuit cal was performed passed and unit working as per manufactures specification.(b)(4).
 
Event Description
The customer reported that while in patient use they were unable to adjust map lower than 20 using the adjust knob.There was no harm to the patient as they were able to continue to use the ventilator.The map was adjusted with the limit knob so there were able to continue using the ventilator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREFUSION
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
jill rittorno
75 n. fairway drive
vernon hills, IL 
8473628056
MDR Report Key5107842
MDR Text Key27062307
Report Number2021710-2015-01777
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Health Professional
Report Date 09/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/17/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-